Signalling: Receiver Quality
CMU-K20/21/22/23/24
1115.6088.12 6.158
E-12
Slot Configuration: Individual (Single-Slot, Circuit Switched)
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:LEVel[:SLOT]:ZERO
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:LEVel[:SLOT]:ONE
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:LEVel[:SLOT]:TWO
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:LEVel[:SLOT]:THRee
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:LEVel[:SLOT]:FOUR
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:LEVel[:SLOT]:FIVE
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:LEVel[:SLOT]:SIX
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:LEVel[:SLOT]:SEVen
<Level>
<Level>
Description of parameters
Def. value
Def. unit
FW vers.
–127 dB to +127 dB
Power of CMU in timeslot no. n
0
dB
V3.05
Description of command
These commands define the levels of the downlink (BS) TCH signal relative to the
Reference Level
set via
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:MSLot:RLEVel
(see p.
6.157
). The
levels are valid for BER tests if the MS is set to multislot operation.
The level range quoted above is restricted by the condition that the absolute level (calculated from the reference
level and the relative individual levels) must not exceed the level ranges of the RF connectors.
Example
With output connector RF2 and a default used timeslot level of –102 dBm, the individual levels can
be set in the range –35 dB to +92 dB, corresponding to absolute levels of –137 dBm to –10 dBm.
Slot Configuration: Individual (Single-Slot, Packet Data)
CONFigure:RXQuality:BER<nr>:CONTrol:PDATa[:TCH]:LEVel[:SLOT]:ZERO
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol:PDATa[:TCH]:LEVel[:SLOT]:ONE
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol:PDATa[:TCH]:LEVel[:SLOT]:TWO
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol:PDATa[:TCH]:LEVel[:SLOT]:THRee
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol:PDATa[:TCH]:LEVel[:SLOT]:FOUR
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol:PDATa[:TCH]:LEVel[:SLOT]:FIVE
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol:PDATa[:TCH]:LEVel[:SLOT]:SIX
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol:PDATa[:TCH]:LEVel[:SLOT]:SEVen
<Level>
<Level>
Description of parameters
Def. value
Def. unit
FW vers.
–127 dB to +127 dB
Power of CMU in timeslot no. n
0
dB
V3.40
Description of command
These commands define the levels of the downlink (BS) TCH signal relative to the
Reference Level
set via
CONFigure:RXQuality:BER<nr>:CONTrol:PDATa:[:TCH]:MSLot:RLEVel
(see p.
6.157
). The levels are
valid for BER tests if the MS is set to multislot operation.
The level range quoted above is restricted by the condition that the absolute level (calculated from the reference
level and the relative individual levels) must not exceed the level ranges of the RF connectors.
Example
With output connector RF2 and a default used timeslot level of –102 dBm, the individual levels can
be set in the range –35 dB to +92 dB, corresponding to absolute levels of –137 dBm to –10 dBm.
CONFigure:RXQuality:BER<nr>:CONTrol:DEFault <Enable>
Default Settings
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
All parameters are set to their default values
Some or all parameters differ from the default values
ON
– V2.00
Description of command
As a
setting command
with the setting
ON
this command sets all parameters of the subsystem to default values
(the setting
OFF
causes an error message).
As a query, this command reads out whether all parameters are set to default values
(ON)
or not
(OFF)
.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...