CMU-K20/21/22/23/24 Connection
Control (Signalling only)
1115.6088.12 6.193
E-12
CONFigure:NETWork[:CSWitched]:AOCHarge:ENABle
<Enable>
Advice of Charge
CONFigure:NETWork[:CSWitched]:AOCHarge
<Value1>, <Value2> ... , <Value7>
<Enable>
Description of parameters
Def. value
Def. unit
ON | OFF
Switch on or off data for advice of charge
ON
–
<Value n> ...
Description of parameters
Def. value
Def. unit
FW vers.
0 to 8191
Values 1 to 7 for calculating the charges
0,0,0,0,0,0,0 –
V1.15
Description of command
Sig. State
The seven numbers denote the following:
Units per interval
Seconds / time interval
Scaling
factor
Unit
increment
Units per data interval
Segments / data interval
Initial secs / t interval
≠
CEST,
Q: all
CONFigure:NETWork[:CSWitched]:SOFFset <
Offset
>
Slot Offset
<Offset>
Description of parameters
Def. value
Def. unit
FW vers.
–7 to +7
Slot Offset
0
–
V3.05
Description of command
Sig. State
This command defines the DL timeslot that the mobile loops back to the uplink main timeslot.
≠
CEST,
Q: all
CONFigure:NETWork:B52Mode <
Mode
>
B52 Mode
<Offset>
Description of parameters
Def. value
Def. unit
FW vers.
SCOD |
MSUP
Speech codec reserved for audio tests
Speech codec supports BLER measurement
SCOD – V3.10
Description of command
Sig. State
This command defines the function of the speech codec.
SOFF,
Q: all
Subsystem NETWork:IDENtity
The subsystem
NETWork:IDENtity
defines the identity of the mobile radio network. The subsystem
corresponds to the table section
Network Identity
in the
Network
tab of the
Connection Control
menu.
CONFigure:NETWork:IDENtity:NCC <
Code
>
NCC
<Code>
Description of parameters
Def. value
Def. unit
FW vers.
0 to 7
Color code of mobile network
0
–
V1.15
Description of command
Sig. State
This command defines the color code of the network
(NCC = network color)
for the signals of the
measuring instrument.
SOFF,
SON
Q: all
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...