Non Signalling: Power Measurements
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.18
E-11
Slot
Count
The hotkey
Slot Count
defines an integer number of timeslots to be measured. The
actual measured time range is larger than the integer number of slots because it
comprises an additional display margin; for details see remote control command
description. The relation between the measured time range and the trigger time is
given by the
Trig. Slot Offset;
see
Fig. 4-3
on p.
4.18
.
The display range is adapted to the
Slot Count
settings by default but can be
modified by means of the
Display Marker – Time Scale
and
Display Marker –
Default Scale
hotkeys. Changing the
Slot Count
overrides the
Time Scale
settings
and restores the default display range.
Remote control
CONFigure:POWer:MSLot:SCOunt
Trig. Slot
Offset
The hotkey
Trig. Slot Offset
defines a delay time (integer number of GSM timeslots)
between the trigger time and the timeslot that is measured in all
Multislot
configurations. In the graphical display, this measured timeslot is marked by
Trg.
Slot Offs. 0
.
•
If
Slot Count
is equal to 1, then the measurement extends over the measured
timeslot plus an appropriate display margin.
•
If
Slot Count
is equal to 2, then the timeslot preceding the measured timeslot
(Trg. Slot Offs. –1)
and the measured slot
(Trg. Slot Offs. 0)
are measured.
•
If
Slot Count
is equal to 3 (4), then
Trg. Slot Offs – 1,
Trg. Slot Offs. 0
and the
next timeslot (the two next timeslots,
Trg. Slot Offs. + 1
and
Trg. Slot Offs. + 2)
are measured.
The beginning of the measured timeslot defines the origin (symbol no. 0) of the
time axis. The measured timeslot is also the reference for the
Timing
measurement; it must be active to obtain valid measurement results.
The relation between the
Trig. Slot Offset
, the
Slot Count
and the measured time
range for a signal with three active timeslots is shown in
Fig. 4-3
below
.
Trigger time
Trig. Slot Offset = 3
Slot Count = 4
Slot Count = 2
Slot Count = 1
Slot Count = 3
Measured time range = default display range
Trg. Slot
Offs. +1
Trg. Slot
Offs. -1
Measured
timeslot:
Trg. Slot
Offs. 0
Trg. Slot
Offs. +2
Fig. 4-3 Trigger slot offset and slot count (for Trig. Slot Offset = 3)
The display range is adapted to the
Slot Count
and
Trig. Slot Offset
settings by
default but can be modified by means of the
Display Marker – Time Scale
and
Display Marker – Default Scale
hotkeys.
Remote control
CONFigure:RFANalyzer:MCONtrol:TSOFfset 0 to 7
Display
Marker
The
Display/Marker
softkey, which is activated by pressing the selected
Marker/Display
softkey again, configures the graphical display.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...