Common: MODulation[:PERRor]
CMU-K20/21/22/23/24
1115.6088.12 6.64
E-12
MODulation[:PERRor]
The subsystem
MODulation[:PERRor][:GMSK]
measures the modulation parameters (frequency and
phase errors) in GMSK modulation. The subsystem corresponds to the measurement menu
Modulation,
application
Ext. Phase Err. GMSK,
and the associated popup menu
Modulation Config
, however, it
does not provide the I/Q Imbalance and the Origin Offset in the I/Q constellation diagram
.
If results for the I/Q Imbalance and the Origin Offset are needed, the slower
MODulation:XPERror
measurement must be used; see p.
6.72
ff.
Note:
GMSK and 8PSK modulation
The
keywords
[:GMSK]
and
:EPSK
in the remote control commands denote GMSK and
8PSK modulation, respectively. The
:EPSK
commands in Signalling measurements are in-
cluded in firmware versions V3.05 and higher. The firmware version numbers quoted in the
command tables refer either to GMSK modulation or EPSK modulation in Non Signalling
mode.
Control of Measurement – Subsystem MODulation[:PERRor][:GMSK]
The subsystem
MODulation[:PERRor][:GMSK]
controls the modulation measurement. It corresponds to
the softkey
Ext. Phase Err. GMSK
in the measurement menu
Modulation
.
INITiate:MODulation[:PERRor][:GMSK]
Start new measurement
⇒
RUN
ABORt:MODulation[:PERRor][:GMSK]
Abort running measurement and switch off
⇒
OFF
STOP:MODulation[:PERRor][:GMSK]
Stop measurement after current stat. cycle
⇒
STOP
CONTinue:MODulation[:PERRor][:GMSK]
Next measurement step (only
stepping mode
)
⇒
RUN
Description of command
FW vers.
These commands have no query form. They start and stop the modulation measurement, setting it
to the status indicated in the top right column.
V1.15
CONFigure:MODulation[:PERRor][:GMSK]:EREPorting <
Mode
>
Event
Reporting
<Mode>
Description of parameters
Def. value
Def. unit
FW vers.
SRQ |
SOPC |
SRSQ |
OFF
Service request
Single operation complete
SRQ and SOPC
No reporting
OFF –
V1.15
Description of command
This command defines the events generated when the measurement is terminated or stopped
(event reporting,
see chapter 5 of CMU manualCMU manual).
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...