Common: SPECtrum:MODulation
CMU-K20/21/22/23/24
1115.6088.12 6.106
E-12
CONFigure:SPECtrum:MODulation[:GMSK]:CONTrol:DEFault
<Enable>
CONFigure:SPECtrum:MODulation:EPSK:CONTrol:DEFault
<Enable>
Default Settings
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
All the parameters of the subsystem are set to default values
At least one parameter of the subsystem differs from its default
value
ON
– V2.00
Description of command
If used as a setting command with the parameter
ON
, this command sets all the parameters of the subsystem to
their default values (
OFF
causes an error message).
In the query format, the command returns
ON
if all the parameters of the subsystem correspond to their default
values, otherwise it returns
OFF
.
Test Configuration
The commands of the following subsystems configure the spectrum due to modulation. They corre-
spond to the
due to Modulation
sections in the
Spectrum Configuration
menu.
Subsystem SPECTrum:MODulation:LIMit:LINE
The subsystem
SPECtrum:MODulation:LIMit:LINE
defines the limit lines, i.e. the tolerance values for
the spectrum due to modulation measurement The subsystem corresponds to the
due to Modulation
sections in the tab
Limit Lines
in the popup menu
Spectrum Configuration.
CONFigure:SPECtrum:MODulation[:GMSK]:LIMit:LINE[:ASYMmetrical]:UPPer<nr>:ENABle
CONFigure:SPECtrum:MODulation:EPSK:LIMit:LINE[:ASYMmetrical]:UPPer<nr>:ENABle
<Enable>
CONFigure:SPECtrum:MODulation[:GMSK]:LIMit:LINE[:ASYMmetrical]:UPPer<nr>
CONFigure:SPECtrum:MODulation:EPSK:LIMit:LINE[:ASYMmetrical]:UPPer<nr>
Limits
<MinPwLevelRel>, <MaxPwLevelRel>, <AbsPwLevel>, <Enable>
Numeric Suffix
Value range
Description of parameters
Def. value
<nr>
1 to11
Measurement point (frequency) no.
Parameters
Value range
Description of parameters
Def. value
<Enable>
<MinPwLevelRel>,
<MaxPwLevelRel>,
<AbsPwLevel>,
<Enable>
ON | OFF
–99.9 dB to 99.9 dB
–99.9 dB to 99.9 dB
–99.9 dBm to 99.9 dBm
ON | OFF
Defined frequency on/off
Limit for relative power below the interpola-
tion range
Limit for relative power above the interpola-
tion range
Alternative absolute power limit
Enable or disable limit check for frequency
point <nr>
ON
See be-
low
See be-
low
See be-
low
Description of command
FW vers.
These commands activate and define limit lines for the spectrum due to modulation measurement.
Each number <nr> denotes a pair of frequency points symmetric to the carrier, <nr>=1 correspond-
ing to
±
0.1 MHz, <nr>=11 to
±
1.8 MHz.
V1.20
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...