Options and Extensions
CMU-K20/21/22/23/24
1115.6088.12 6.216
E-12
PROCedure:SIGNalling:PDATa[:TCH]:MSLot:FHOPping:ENABle
<Enable>
Enable Frequency Hopping
<Sequence>
Description of parameters
Def. value
Def. unit
FW vers.
ON | OFF
Enable or disable frequency hopping
OFF
–
V3.40
Description of command
Sig. State
This command enables or disables frequency hopping in the downlink traffic channel.. The hopping
sequences are defined via
CONFigure:BSSignal:PDATa[:TCH]:MSLot:FHOPping:SEQuence
; see p.
6.217
.
CEST
Q: all
Subsystem MSSignal:PDATa (RF Signal of MS under Test)
The subsystem
MSSignal:PDATa
configures the RF signal that the MS under test is to transmit in
GPRS test mode. It corresponds to the
Packet Data
section in the
MS Signal
tab of the popup menu
Connect. Control
.
CONFigure:MSSignal:PDATa[:TCH]:MSLot:SCONfig
Uplink Slot Configuration
<Enable_0>,..., <Enable_7>, <Gamma_0>,... <Gamma_7>
<Enable_n>
Description of parameters
Def. value
Def. unit
ON | OFF
Enable or disable uplink timeslot no. n
n = 3: ON
–
<Gamma_n>
Description of parameters
Def. value
Def. unit
FW vers.
0 to 31
Power control parameter
Γ
CH
in timeslot no. n
3
– V3.05
Description of command
Sig. State
This command defines the slot configuration and the channel-specific power control parameters
Γ
CH
that the MS is to use in test mode (default parameters). In the default setting, In the default setting,
only slot 3 is enabled. Slot no. 3 is also the main timeslot; see
CONFigure:BSSignal:PDATa[:TCH]:MSLot:MTIMeslot
command on p.
6.217
.
≠
TEST,
Q: all
Subsystem BSSignal:PDATa (RF Signal of Base Station/CMU)
The subsystem
BSSignal:PDATa
configures the RF signal that the CMU transmits in packet data
transfer mode. It corresponds to the
Packet Data
section in the
BS Signal
tab of the popup menu
Connect. Control
.
CONFigure:BSSignal:PDATa[:TCH]:MSLot:PZERo
<P0>
P0
Parameter
<P0>
Description of parameter
Def. value
Def. unit
FW vers.
0 dB to 31 dB
Value of P0
5
dB
V3.05
Description of command
Sig. State
This command defines the downlink power control parameter P0.
≠
TEST,
Q: all
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...