CMU-K20/21/22/23/24
Options and Extensions
1115.6088.12 6.225
E-12
CONFigure:NETWork[:CSWitched]:AMR:HRATe:RSETting
AMR Rate Set, Half Rate
<CM4>, <CM3>, CM2>, <CM1>,
<ThrDown4>, <ThrUp3>, <ThrDown3>, <ThrUp2>,
<ThrDown2>, <ThrUp1>
<CM4>, <CM3>, <CM2>, CM1>
Description of parameters
Def. value
Def. unit
FW vers.
C1220 | C1020 | C0795 |
C0740 | C0670 | C0590 |
C0515 | C0475 |
OFF
User bit rate for codec modes 4 to 1.
Mode switched off
C0795,
C0670,
C0590,
C0515,
–
<ThrDown4>
Description of parameters
Def. value
Def. unit
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Lower decision threshold for switching
between modes 4 and 3
15.0, dB
<ThrUp3>
Description of parameters
Def. value
Def. unit
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Upper decision threshold for switching
between modes 3 and 4
17.0, dB
<ThrDown3>
Description of parameters
Def. value
Def. unit
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Lower decision threshold for switching
between modes 3 and 2
12.5, dB
<ThrUp2>
Description of parameters
Def. value
Def. unit
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Upper decision threshold for switching
between modes 2 and 3
15.0, dB
<ThrDown2>
Description of parameters
Def. value
Def. unit
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Lower decision threshold for switching
between modes 2 and 1
11.0, dB
<ThrUp1>
Description of parameters
Def. value
Def. unit
FW vers.
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Upper decision threshold for switching
between modes 1 and 2
13.0 dB V3.40
Description of command
Sig. State
These commands select four codec modes and define the decision thresholds for changing the codec
mode. The settings are valid for Half Rate AMR speech coder tests; see command
CONFigure:NETWork[:CSWitched]:SMODe:TRAFfic
on p.
6.197
. The instrument rejects the
settings (SCPI error –221, Settings conflict) unless the values meet all of the following conditions:
•
The rates must be in descending order so that <CM4> is the largest bit rate.
•
Up to 3 codec modes can be switched off. OFF must be the first values of the parameter list,
preceding the used codec modes.
•
Thresholds must be in descending order so that <ThrDown2>
≤
<ThrDown3>
≤
<ThrDown4> and
<ThrUp1>
≤
<ThrUp2>
≤
<ThrUp3>.
•
The hysteresis must be positive so that Up (j)
≥
Down (j + 1) for j= 1 to 3
To query the DL codec mode requested by the MS use
[SENSe:]MSSinfo:AMR:HRATe:DLCMode?
and
[SENSe:]MSSinfo:AMR:FRATe:DLCMode?
(see p.
6.209
).
≠
CEST,
Q: all
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...