CMU-K20/K21/K22/K23/K24
Connection Control (Non Signalling only)
1115.6088.12 6.9
E-12
TRIGger[:SEQuence]:THReshold:RFPower
<Threshold>
Level – RF Power
<Threshold>
Parameter description
Def. value
Default unit
FW vers.
LOW |
MEDium |
HIGH
Low trigger threshold
(RF Max. Level
– 26 dB)
Medium trigger threshold
(RF Max. Level
– 16 dB)
High trigger threshold
(RF Max. Level
– 6 dB)
MEDium –
V3.10
Command description
This command sets the RF input signal level at which the measurement is triggered relative to the maximum RF
input level; see
[SENSe:]LEVel:MAXimum
. The setting has effect for trigger source
RFPower
only (see
TRIG:SEQ:SOUR
).
TRIGger[:SEQuence]:THReshold:IFPower
<Threshold>
Level – IF Power
<Threshold>
Parameter description
Def. value
Default unit
FW vers.
–47 dB to 0 dB
IF power threshold
–26
dB
V3.10
Command description
This command sets the IF signal level at which the measurement is triggered. The IF power threshold is defined
relative to the maximum RF input level; see
[SENSe:]LEVel:MAXimum
. The setting has effect for trigger source
IFPower
only (see
TRIG:SEQ:SOUR
).
TRIGger[:SEQuence]:SLOPe
<Slope>
Slope
<Slope>
Parameter description
Def. value
Default unit
FW vers.
POSitive |
NEGative
Rising edge
Falling edge
POS –
V3.10
Command description
This command qualifies whether the trigger event occurs on the
Rising Edge
or on the
Falling Edge
of the trigger
signal. The setting has no influence on
Free Run
measurements (see
TRIG:SEQ:SOUR
).
TRIGger[:SEQuence]:SOURce:EXTernal
<Source>
Ext. Trigger (AUX 3/4)
<Source>
Description of parameters
Def. value
Def. unit
FW vers.
PIN6 | PIN7 | PIN8
Pin for external trigger signal
PIN8
–
V3.10
Description of command
This command determines the pins on the AUX 3 or AUX4 connectors used for the external trigger signal. The
setting only has effect if the trigger source is an
External
signal.
TRIGger[:SEQuence]:DEFault <Enable>
Default Settings
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
The parameters are set to their default values
Some or all parameters differ from the default values
ON
– V1.15
Description of command
If used as a setting command with the parameter
ON
this command sets all parameters of the subsystem to their
default values (the setting
OFF
causes an error message).
If used as a query the command returns whether all parameters are set to their default values
(ON)
or not
(OFF)
.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...