CMU-K20/-K21/-K22/-K23/-K24
Non Signalling Mode
1115.6088.12 2.9
E-10
Additional Information...
Alternative Settings
and Measurements
... on Step 2
"
Power Configuration menu
The
Power Configuration
menu contains three tabs defining
•
Measurement control and statistical settings
(Control),
•
The tolerance template for the burst
(Limit Lines),
•
PCL-dependent limits for the average burst p
ower (Lim-
its)
'
chapter
3.
Settings made in the
Power Con-
figuration
menu apply to power
measurements only.
Settings made in the
Connect.
Control
menus apply to the entire
function group and mode
GSM900-MS Non Signalling
.
#
Repetition mode and Stop Condition
If no stop condition is imposed
(Stop Condition = None),
the
Repetition
mode determines whether the measurement is
•
Continued
until
explicitly stopped by the operator
(Con-
tinuous),
•
Stopped after one statistic count
(Single Shot).
By default, a statistic count comprises 100 bursts. With
Stop
Condition = On Limit Failure
, the measurement is stopped af-
ter the first burst which is out of tolerance.
'
chapter
3.
The
Statistic Count
is defined in
the
Control
tab of the
Power Con-
figuration
menu.
The stop condition
On Limit Fail-
ure
should be selected if the limit
check represents the main pur-
pose of the measurement.
The limits can be modified in the
Limit Lines
tab of the
Power Con-
figuration
menu.
$
Measurement in the HLT state
The average and peak power of the last burst measured is
indicated in the output fields
Average
and
Peak
.
In contrast, the modulation measurement is still running. The
results for the frequency and phase errors are periodically
updated.
'
CMU
manual
See the sections on measure-
ment control in chapter 3 and 5.
... on Step 3
%
Power menu
The diagram in the
Power
menu, application
Power/t Norm.
GMSK
shows a normal burst with a length of 148 bits (plus a
guard period of 8.25 bits). The time scale of the diagram
ranges from –10 bits to 156¾ bits covering the useful part,
the rising and falling edges of the burst. The ordinate ranges
from –80 dB to +10 dB, the 0-dB reference level is equal to
the carrier power.
Note that settings made previously (
Power Configuration
menu) are preserved in the whole measurement group. Ac-
cordingly, the status of the measurement is still
HLT.
The dia-
gram is fixed showing the last burst measured.
'
chapter 4, p. 4.9 ff.
The GSM power template is de-
fined relative to the carrier power.
For low signal powers, a looser
absolute limit is to be applied at
the beginning and the end of the
power ramp (areas 1, 2, 7, and
8). This yields the distorted tem-
plate that we observe in the pre-
sent example.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...