CMU-K20/K21/K22/K23/K24
Measurement Groups (Non Signalling only)
1115.6088.12 6.17
E-12
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer<AreaNr>:DYNamic<RangeNr>:ENABle
CONFigure:POWer[:NORMal]:EPSK:LIMit:LINE:UPPer<AreaNr>:DYNamic<RangeNr>:ENABle
<Enable>
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer<AreaNr>:DYNamic<RangeNr>
CONFigure:POWer[:NORMal]:EPSK:LIMit:LINE:UPPer<AreaNr>:DYNamic<RangeNr>
<fromTPCL>,
<toTPCL>, <Correction>, <Enable>
Dynamic Correction
<fromTPCL>
Description of parameters
Def. value
Def. unit
0 to 31 |
OFF
First template PCL for which area <AreaNr> is
changed
See table below
–
<toTPCL>
Description of parameters
Def. value
Def. unit
0 to 31 | OFF
Last template PCL for which area <AreaNr> is
changed.
See table below
–
<
Correction>
Description of parameters
Def. value
Def. unit
-10 dB to
+10 dB | OFF
Correction value (relative) for the limit line in area
<AreaNr> applied for all template PCLs between
<fromTPCL> and <toTPCL>
See table below
dB
<
Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON | OFF
Enable or disable dynamic correction in the cur-
rent limit line area and PCL range
See table below
–
V2.00
Description of command
These command activates and defines dynamic correction of the upper limit line of area <AreaNr> (<AreaNr> = 1
to 16) depending on the template PCL range <RangeNr> (<RangeNr> = 1 to 10 for each area). The template
PCL is set via
CONFigure:RFANalyzer:TPCL
.
In the areas no. 3 and 6, the following ranges are defined (both modulation schemes):
Range fromTPCL
toTPCL
Correction
Enable
1 16 16 +2.0
dB
ON
2 17 17 +4.0
dB
ON
3 18 19 +5.0
dB
ON
4
OFF OFF 0.0
dB
OFF
5
OFF OFF 0.0
dB
OFF
6
OFF OFF 0.0
dB
OFF
7
OFF OFF 0.0
dB
OFF
8
OFF OFF 0.0
dB
OFF
9
OFF OFF 0.0
dB
OFF
10 OFF OFF 0.0
dB
OFF
In the remaining areas, the dynamic limit line correction is disabled in all ranges.
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer<AreaNr>:ALL:DYNamic:ENABle
CONFigure:POWer[:NORMal]:EPSK:LIMit:LINE:UPPer<AreaNr>:ALL:DYNamic:ENABle
<Enable>
Dynamic Correction on/off
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON | OFF
Switch dynamic correction on or off
ON
–
V2.00
Description of command
This command switches the dynamic correction of the upper limit area <nr> for all ten TPCL ranges on or off.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...