CMU-K20/K21/K22/K23/K24
Signalling: Power Measurements
1115.6088.12 4.103
E-12
Slot
Count
The hotkey
Slot Count
defines an integer number of timeslots to be measured. The
actual measured time range is larger than the integer number of slots because it
comprises an additional display margin; for details see remote control command
description. The
Meas. Slot
hotkey defines the position of the measurement range
within the TDMA frame
;
see
Fig. 4-81
on p.
4.185
.
The display range is adapted to the
Slot Count
settings by default but can be
modified by means of the
Display Marker – Time Scale
and
Display Marker –
Default Scale
hotkeys. Changing the
Slot Count
overrides the
Time Scale
settings
and restores the default display range.
Remote control
CONFigure:POWer:MSLot:SCOunt
Meas
Slot
The hotkey
Meas. Slot
determines the timeslot that is measured in all Multislot
configurations. In the graphical display, this measured timeslot is marked by Meas.
Slot.
Meas. Slot is a general parameter that is valid for all measurement groups and also
provided in the
Analyzer
tab of the
Connection Control
menu. For more information
see p.
4.185
.
Remote control
CONFigure:MCONtrol:MSLot:MESLot
c) P/Frame, P/Slot Graph, P/Slot Table
The settings accessible via the measurement control softkey, the
Application, Analyzer Level, MS
Signal,
and BS Signal
softkey are identical with those of application
P/t Normal GMSK
, see above. The
Marker/Display
softkey is not needed.
d) P/PCL
The settings of the
Application, Analyzer Level
, and
BS Signal
softkey are identical with those of
application
P/t Normal GMSK
, see above. In application P/PCL, the
P/PCL – Channel Count
and the
MS Signal
softkey are used for channel selection.
P/PCL
The
P/PCL
softkey controls the power measurement and indicates its status (RUN |
HLT | OFF).
Channel
Count
The
Channel Count
hotkey determines whether the P/PCL measurement is
performed on 3 or 7 different GSM channels. If only 3 channels are selected, the
P/PCL measurement is faster; the last four columns in the output table of the
Power
menu are suppressed.
Remote control
CONFigure:POWer:PCL:CCOunt C3 | C7
MS
Signal
The
MS Signal
softkey defines seven channels for the P/PCL measurement.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...