Connection Control (Signalling only) CMU-K20/21/22/23/24
1115.6088.12 6.190
E-12
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:MTIMeslot
<Slot_No>
Main Timeslot
<Slot_No>
Description of parameters
Def. value
Def. unit
FW vers.
0 to 7
Main timeslot, used for signalling
3
–
V3.05
Description of command
Sig. State
This command defines the timeslot that the MS and the BS/CMU use for signalling (default value).
Changing the main timeslot also overwrites the
Meas.- Slot
(command
CONFigure:MCONtrol:MSLot:MESLot
). If used in the CEST state, this command overwrites the
main timeslot set via
PROCedure:SIGNalling[:CSWitched][:TCH]:MSLot:SCONfig
(see p.
6.181
).
≠
CEST,
Q: all
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:LMODe
<Mode>
Level Mode
<Slot_No>
Description of parameters
Def. value
Def. unit
FW vers.
UUN |
IND
Used timeslot and unused timeslot levels
Individual timeslot levels
UUN –
V3.05
Description of command
Sig. State
This command determines whether the CMU uses the used/unused timeslot level scheme (two
different levels) or individual levels in all timeslots.
≠
CEST,
Q: all
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:SCONfig:UUNused
Slot Configuration: Used/Unused
<Enable_0>,<Enable_1>, ..., <Enable_7>
<Enable_n>
Description of parameters
Def. value
Def. unit
FW vers.
ON | OFF
Enable or disable timeslot no. n
see below
–
V3.05
Description of command
Sig. State
This command defines the active timeslots slots of the BS signal (default values) and implicitly
defines the levels for the
Used/Unused
level mode (see command
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:LMODe
on p.
6.190
).
•
The level in all enabled timeslots is given by the
Used Timeslot Level
defined via
CONFigure:BSSignal[:CSWitched][:TCH]:LEVel:UNTimeslot
(see p.
6.188
).
•
The level in all disabled timeslots is given by the
Unused Timeslot Level
def. via
CONFigure:BSSignal[:CSWitched][:TCH]:LEVel:UTIMeslot
(see p.
6.187
).
In the default setting, timeslots 3 and 4 are enabled. By default, slot no. 3 is also the main timeslot;
see command
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:MTIMeslot
on p.
6.190
.
Note:
The <Enable_n> parameters also apply in individual level mode; they overwrite the
parameters in the
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:SCONfig:INDividual
command.
≠
CEST,
Q: all
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...