CMU-K20/21/22/23/24 Signalling: POWer:PCL
1115.6088.12 6.143
E-12
Subsystem POWer:PCL:CONTrol
CONFigure:POWer:PCL:CONTrol:REPetition <
Repetition
>
,<StopCondition>,<Stepmode>
Test Cycles
<Repetition>
Description of parameters
Def. value
Def. unit
CONTinuous |
SINGleshot |
1 to 10000
Continuous measurement
(
continuous
, until
STOP
or
ABORT)
Single measurement
(single shot
, until
Status = RDY)
Multiple measurement
(
counting
, until
Status = STEP | RDY)
SING –
<StopCond>
Description of parameters
Def. value
Def. unit
SONerror |
NONE
Stop measurement in case of error
(stop on error)
Continue measurement even in case of error
NONE –
<Stepmode>
Description of parameters
Def. value
Def. unit
FW vers.
STEP |
NONE
Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode
NONE –
VV2.00
Description of command
This command determines the number of statistics cycles and the stepping mode for the measurement.
Note:
In the case of READ commands (
READ:
...), the <Repetition> parameter has no effect; the
measurement is always stopped after a single shot.
Subsystem POWer:PCL:CHANnel
The subsystem
POWer:PCL:CHANnel
defines three or seven channels for the
P/PCL
measurement.
The subsystem corresponds to the
Channel Count
hotkey and the MS Signal softkey in the
measurement menu
P/PCL.
CONFigure:POWer:PCL:CCOunt <Channels>
Channel
Count
<Channels>
Description of parameters
Def. value
Def. unit
FW vers.
C3 | C7
3 or 7 different channels measured
C3
–
V2.15
Description of command
This command defines whether three or seven different channels are measured in the P/PCL measurement.
CONFigure:POWer:PCL:CHANnel <Channel1>, ..., <Channeln>
Channel
Channel<nr>
Value range
Def. value: Cannel 1,2,3
Def. unit
FW vers.
GSM400
GSM850
GSM900
GSM1800
GSM1900
259 to 293, 306 to 340
128 to 251
0 to 124; 955 to 1023
512 to 885
512 to 810
259,276,293
128,190,251
1,62,124
512,698,885
512,661,810
– VV2.00
4
4
In firmware versions <2.15, only three channels could be measured.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...