CMU-K20/21/22/23/24 Common: POWer:XSLot
1115.6088.12 6.47
E-12
CONFigure:POWer:XSLot:SCOunt
<Group>[,<Number>]
Slot Count
<Group>
Description of parameters
Def. value
Def. unit
S128 | S256 | S384 | S512
Number of slots measured, if an integer
multiple of 128
S128 –
<Number>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 512
Number of slots measured, for FW vers.
V3.05 and higher
128 –
V2.15
Description of command
This command defines the total number of slots measured. The first parameter sets the slot count in multiples of
128 (Sn where n = 1 to 4) and defines the number of returned values. The second parameter is equal to the
Slot
Count
in manual control and must be used if a
Slot Count
m
≠
n*128 is desired. m supersedes n such that n
≥
m.
Examples:
•
CONF:POW:XSL:SCO S128,128
causes 128 values to be calculated and returned.
CONF:POW:XSL:SCO S128,129
is equivalent to
CONF:POW:XSL:SCO S256,129
. 129 values are measured.
The output arrays of the
XSLot
measurement group consist of 129 valid and 127 invalid (NAN) results.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...