Signalling: POWer[:NORMal] CMU-K20/21/22/23/24
1115.6088.12 6.120
E-12
Measurement Groups (Signalling only)
The measurement groups in this section are either provided in
Signalling
mode only or implemented
with major differences in the two test modes.
POWer[:NORMal]
The subsystem
POWer[:NORMal]
measures the MS transmitter output power versus time for normal
bursts. The subsystem corresponds to the measurement menu
Power,
application
P/t Normal...,
and
the associated popup menu
Power Configuration.
Note1: Measurements and signalling states
To perform any kind of measurement and obtain a meaningful result, an appropriate test setup
is required (see application examples in chapter 2 of this manual). Consequently, for the
measurements reported in the following sections, the Call Established (CEST) signalling state
must be reached before any of the commands retrieving test results
(READ...?,
FETCh...?
,
SAMPle...?
, or
CALCulate...LIMit?)
can be used. Test configurations,
however, can be defined any time.
Note2: GMSK and 8PSK modulation
T
he keywords
[:GMSK]
and
:EPSK
in the remote control commands denote GMSK and 8PSK
modulation, respectively. The
:EPSK
commands in Signalling measurements are included in
firmware versions V3.05 and higher. The firmware version numbers quoted in the command
tables refer to GMSK modulation.
Control of Measurement – Subsystem Power[:NORMal]
The subsystem
POWer[:NORMal]
controls the normal burst power measurement.
INITiate:POWer[:NORMal][:GMSK]
INITiate:POWer[:NORMal]:EPSK
Start new measurement
⇒
RUN
ABORt:POWer[:NORMal][:GMSK]
ABORt:POWer[:NORMal]:EPSK
Abort running measurement and switch off
⇒
OFF
STOP:POWer[:NORMal][:GMSK]
STOP:POWer[:NORMal]:EPSK
Stop measurement after current stat. cycle
⇒
STOP
CONTinue:POWer[:NORMal][:GMSK]
CONTinue:POWer[:NORMal]:EPSK
Next measurement step (only stepping mode)
⇒
RUN
Description of command
FW vers.
These commands have no query form. They start or stop the measurement, setting it to the status
indicated in the top right column.
V1.15
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...