Passport V Operating Instructions
5 - 1
5.0
Respiration Monitoring
5.1
Description
Intended patient types: Adult, Pediatric, Neonate.
Respirations, or the amount of breaths per minute, are measured by two methods in the
Passport V
:
The first method is thoracic impedance through the ECG signal. The respiration signal is
measured between two ECG electrodes:
• RA and LA of ECG Lead I, or
• RA and LL of ECG Lead II
The second is by CO
2
exchange via internal CO
2
or via the external Gas Module. Section
5.3.1 discusses the monitoring of CO
2
with thoracic impedance. Chapter 10.0 discusses the
monitoring of CO
2
with Sidestream and Microstream capnography.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...