Passport V Operating Instructions
7 - 5
NIBP Monitoring
NIBP Menu
7.3
NIBP Menu
To display the
NIBP Menu
:
1.
On the front panel:
Press the
Normal Screen
key to return to the normal screen.
2.
On the normal screen:
Select
Parameters
〉
NIBP
or
Select the
NIBP
tile
FIGURE 7-3
NIBP Menu
NIBP Menu
MENU ITEM
SELECTIONS
COMMENTS
Normal Screen
—
Select to return to normal screen (or press the
Normal Screen
key on the front panel).
Set Start Pressure
Adult: 100 to 280
(default = 180)
Pediatric: 60 to 180
(default = 140)
Neonate: 40 to 120
(default = 100)
Select to adjust initial NIBP pump up pressure in
increments of 5. The selection range varies
according to the Patient Size.
The Set Start Pressure cannot be changed during
an NIBP measurement or the five minutes STAT
measurement.