Arrhythmia Analysis (Optional)
ECG Monitoring
4 - 28
Passport V Operating Instructions
4.8
Arrhythmia Analysis (Optional)
NOTE:
Arrhythmia analysis is available for Adult and Pediatric
patients only.
WARNING: Due to physiologic differences in the patient population, the
The
Passport V
is capable of identifying ventricular arrhythmia patterns in Adult and
Pediatric size patients. Arrhythmia analysis may be enabled or disabled via the
Arrhythmia Menu
. By default, arrhythmia analysis is enabled if the option is installed.
FIGURE 4-26
Arrhythmia Menu
Arrhythmia alarm calls are classified as Priority 1 or Priority 2.
Asystole, Ventricular Tachycardia, and Ventricular Fibrillation are classified as Priority 1
and the priority level cannot be changed by the user. In addition, these alarms will
sound continuously until the user presses the
Silence
or
Silence All
key, regardless of
whether the patient’s condition has improved.
The other arrhythmia alarms (listed below) are classified as Priority 2 by default. The
characteristics and priority level of all non-lethal alarms can be changed at the user’s
discretion via the
Arrhythmia Setup
menu option.
The following alarm calls can be made when Arrhythmia Analysis is set to “All On” (default
setting):
• Asystole, Ventricular Tachycardia, Ventricular Fibrillation, Ventricular Rhythm, Run,
PVC/Min, Couplet, Bigeminy, Trigeminy, Irregular HR and Bradycardia.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...