Prompt Messages
Messages
14 - 18
Passport V Operating Instructions
14.3.9
Main Control
System
GM: Unknown Agent
Appears when the system detects
a gas that does not match the
spectroscopic signatures of the
five known anesthetic agents.
Use recognized agent.
Sampling Error
Appears when a sampling error
occurs on one or more Gas
Module channels during
calibration.
Repeat calibration
procedure. If problem
persists, contact Technical
Support.
Not Ready For Calibration
Appears when the Gas Module
is unable to initialize calibration.
Repeat calibration
procedure. If problem
persists, contact Technical
Support.
Zeroing Error
Appears when the Gas Module
cannot perform a
Zeroing
during calibration.
Repeat calibration
procedure. If problem
persists, contact Technical
Support.
GM: Communication Error
System can't communicate
correctly with Resp module,and
receive the error data packets.
Restart the monitor.If the error
still appears,please contact
Technical Support.
GM: Disconnected
Appears when the
Passport V
cannot detect signals being sent
by the Gas Module.
Ensure Gas Module is turned
on and interface cable is
properly connected. If problem
persists, contact Technical
Support.
ALARM MESSAGE
REASON
SOLUTION
Power Board Communication
Error
Abnormal power board
communication.
Contact Technical Support.
Power Board Initialization
Error
Contact Technical Support.
Power Board SelfTest Error
Contact Technical Support.
Power Board Communication
Stop
Contact Technical Support.
RT Clock Need Reset
The real time clock need be
reset, as the data read out is the
chip’s default value.
Contact Technical Support.
RT Clock Not Exist
The data written in the real time
clock chip is inconsistent with the
data read out.
Contact Technical Support.
Host Temp Too High
Internal temperature of the host
is too high. This may result from
too high room temperature,
abnormal fan heat dissipation,
or abnormal internal circuits.
Turn off the host until its internal
temperature reduces to the safe
temperature range.
Data Card Error
Data storage card abnormal
Contact Technical Support.
Cooling Fan Failure
The unit running on AC power
and the cooling fan stop work.
Contact Technical Support.
Keyboard Communication
Error
Contact Technical Support.
Keyboard Communication
Stop
Contact Technical Support.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...