Print Setup Menu
Printing
16 - 2
Passport V Operating Instructions
Waveforms may be printed at the time of an alarm by selecting
Active
in response to
On Alarm
in the
Alarm Setup Menu
, which can be accessed by pressing the
Alarms:
Limits
key. Waveforms may also be printed at regular intervals by selecting
Interval
. Available selections are
OFF, 1 min, 5 min, 10 min, 15 min, 20 min, 30
min, 1 hr,
and
2 hr
.
To choose which waveform will print to the local printer select
Waveform 1, 2, or 3
and
then select any of the available choices.
The following printer options may be available:
Local
,
Remote, and Local/Remote
. A
default printer or combination of printers can be set.
Local
printing will initiate a printout delivered via the local printer.
Remote
printing will send a printout to a remote printer (HP1505N, HP4015N) connected
to the
Passport V
.
During printing, if the operator changes a setting (e.g., scale, lead name), the report will
mark the change at the changed time.
16.2.1
Local Printer (Optional)
The
Passport V
local printer can provide a printed record of all patient parameters. It is a
three-trace thermal strip chart local printer with an integral paper spool. The local printer uses
plain white thermal paper, 5 cm wide.
NOTE:
All grid patterns and data are printed by the local printer.
Operation of the Local Printer
1.
Select the waveforms to print to the local printer via the
Print Setup Menu
using the
Navigator Knob. To change the waveforms that appear, select
Waveform 1, 2, or 3
and choose from the available parameters.
2.
To print, select a printer from
Print Setup Menu
〉
Select Printer
. The printer
selection is
Local
by default.
3.
Press the
STRIP
key to initiate printing.
4.
Press the
STRIP
key again to abort the printout. Select printouts will also require the use
of alternate menus or keys.
During a waveform printout, tic marks appear at one second intervals, for reference.
Printouts from the Local Printer
Strip Printing
When the
STRIP
key is pressed a 16 or 32 second strip is printed, acccording to the “Print
Length” setting. The source of the printout is indicated on the strip as
REALTIME REPORT
.
Continuous ECG Printing
If a real-time, continuous printing of ECG data is required, press the
Cont ECG
key. To abort
printing, press the
Cont ECG
,
Strip
, or
Print Trend
key.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...