Passport V Operating Instructions
14 - 11
Messages
Prompt Messages
14.3.5
NIBP Prompt Message
ALARM MESSAGE
REASON
SOLUTION
NIBP: Self Test Error
Failed self-test.
Sensor or A/D sampling may
have an error.
Power cycle unit. If message
reappears, contact Technical
Support.
NIBP: Communication Error
NIBP module communicatuon
error, the command can not be
sent correctly.
Power cycle unit. If message
reappears, contact Technical
Support.
NIBP: System Error
System error. After start-up, the
inflating pump, A/D sampling
unit, pressure sensor have an
error or the pointer has an error
when the software is in running
process.
Power supply is not stable or
circuit failure leads to voltage
error.
Power cycle unit. If message
reappears, contact Technical
Support.
NIBP: Cuff Overpressure
The hardware overpressure limit
has been exceeded.
Overpressure. The cuff pressure
exceeds 297 mmHg in ADU
mode, 240 mmHg in PED mode
or 147 mmHg in NEO mode.
Power cycle unit.
If message reappears, contact
Technical Support.
Pneumatic Leak
Leakage. In Pneumatic check, air
leakage is found in hose.
Change the cuff.
NIBP: Unable to Measure
Unable to make measurement
after three automatic retries.
Check Patient.
Retry measurement. If message
reappears, power cycle unit.
If message reappears, contact
Technical Support.
NIBP: Resetting...
/
NIBP: Reset Failed
Reset has failed.
Power cycle unit.
If message reappears, contact
Technical Support.
NIBP: Idle
Appears when NIBP
measurement is completed and
enters the wait status.
This is normal operation.
Press 'Start' key to start a
measurement, the message is
cleared.
NIBP: Press the START key
Appears when NIBP Interval is
set to anything other than Off or
STAT.
This is normal operation.
Press 'START' key to start
automatic measurement.
NIBP: Manual Mode
Displayed while system is idle.
Note: This is not displayed while
in the interval mode.
Press
START
to take a single
measurement. Select an
interval and start timed
measurements.
NIBP: Interval
Displayed during the interval
between two timed
measurements.
Press
STOP
to suspend timed
measurements.
Change timer to
OFF
to stop
timer.
NIBP: Measuring
Displayed during a
measurement.
Cuff pressure is also displayed.
Press
STOP
to suspend a
measurement and deflate the
cuff.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...