Messages
Prompt Messages
GM: O
2
Zero Error
Appears when the system has
been unable to successfully zero
the O
2
sensor.
Manually start zeroing the
system again. If problem
persists, contact Mindray
Technical Support.
GM: Cannot Zero...Retrying
Appears when the
Passport V
requests
Zeroing
(either on the
automatic cycle or by a user
request) and the Gas Module is
unable to initialize the cycle.
Allow system to retry without
intervention. If problem persist,
contact Mindray Technical
Support
GM: Zero In Progress
Appears when the system is
zeroing all of its channels. This
appears whether initiated by the
user or is automatic.
This is normal operation.
Wait for message to clear.
GM: Warming Up
Appears when the system has
been turned on, and the sensors
have not reached their stable
operating temperature.
Wait for the message to go
away. It takes up to five
minutes for the device to warm
up.
GM: Air Leak
Appears when the system detects
a pneumatic leak.
Also may appear when the Gas
Module has been turned on
without a sample line attached.
Gas Module has been on for a
long period of time without the
Passport V
monitor being on.
Turn Gas Module and
Passport V
Off.
Install/check sample lines,
filters, water trap and electrical
connections.
Turn off Gas Module.Turn on
Gas Module and
Passport V
monitor
GM: Replace Trap
Indicates residue build-up on the
water trap membrane that is
decreasing air flow.
Replace water trap reservoir.
GM: Exhaust Blocked
Appears when the system detects
a blockage at the exhaust gas
outlet, as indicated by an
increase in internal pressure.
Remove waste gas
scavenging assembly, check if
message disappears.
Check exhaust line for
blockage and clear if possible.
If message persists contact
Mindray Technical Support.
GM: Failed
Appears when the Gas Module
detects an unrecoverable error in
its own operation.
Contact Mindray Technical
Support.
GM: Occlusion
Appears when the system detects
an obstruction in the sampling
line or the water trap bottle is
full.
Empty and rinse water trap.
Change water trap if
necessary. Check sampling line
and filter for blockage,
clear sampling line if possible.
Replace sampling line and/or
filter if necessary.
Check exhaust line for
blockage and clear if possible.
If problem persists, contact
Mindray Technical Support.
GM: Pump Off
Appears when the system has
turned off the pump due to a
pneumatic error.
Restart the pump from the
Gas
Menu
. If problem persists,
contact Mindray Technical
Support.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...