Passport V Operating Instructions
19 - 23
Appendix
Performance Characteristics
after entering measurement mode, periodically for durations of typically 10 seconds at
a rate of at most once per hour.
• If a change of 8 °C from the last AZ is detected.
• If a pressure change of 20 mmHg relative to the last ambient pressure measurement
(less than the purge threshold) for a period of 30 seconds is detected. The module will
be able to detect a real change in the ambient pressure and a pressure change due to
partial blockage of the FilterLine.
The module prevents the triggering of an SFM in the following situations:
• In case of purging until the end of this state.
• During a breath absence period which follows a valid breath.
• While waiting a minimum of 20 seconds for host SFM enable command. (After the 20-
second opportunity given to the host to schedule the SFM passes, the module
schedules the SFM according to a priority determined by current conditions).
DPM CO
2
System Response Time:
The system response time (with a standard-length FilterLine)
which includes the delay time and rise time (10% to 90%) in
response to a step change in the CO
2
concentration is 2.9
seconds typical.
Rise Time (Adult and
Neonatal):
190 msec maximum
Delay Time:
2.7 seconds typical
Pump Calibration Interval:
No routine calibration is required. The module should initially
be calibrated after 1200 operating hours, then once a year
or after 4,000 operating hours, whichever comes first.
Initialization Time:
Time to full accuracy (power on to first reading): Typically 30
seconds for both reading and waveform (max 180 seconds).
At full accuracy when value appears.
CO
2
CO
2
measurement range:
0-99 mmHg
Measurement accuracy:
±2mmHg (0-40mmHg)
±5% of reading (41-76mmHg)
±10% of reading (77-99mmHg)
Resolution:
1mmHg
Sampling rate:
50Hz
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...