Passport V Operating Instructions
xi
Warnings
Introduction
WARNING: Use of the Passport V in the vicinity of explosive anesthetics
and in the presence of electromagnetic interference or power
overloads caused by electrosurgical or diathermy instruments
could create a potential hazard or could damage the monitor.
WARNING: The use of gas sampling accessories in Gas Module 3 other than
those specified may cause significant measurement errors and
patient risk.
WARNING: Use of accessories, transducers and cables other than those
specified in the manual may result in increased Electromagnetic
Emissions or decreased Electromagnetic Immunity of the Gas
Module 3.
WARNING: With the exception of stacking on a Gas Module with the
appropriate mounting brackets, the Passport V should not be
used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the Passport V and Gas Module
should be observed to verify normal operation in the
configuration in which they will be used.
WARNING: If the water trap breaks or becomes damaged during
operation, there is a risk that bacteria and/or mucus may
contaminate the Gas Module.
WARNING: The Gas Module must not be used with flammable anesthetic
agents.
WARNING: SpO
2
sensors, SpO
2
accessories, and temperature probes
should be disposed of in accordance with local regulations.
WARNING: Shut down the monitor and disconnect all power cords from the
outlet before cleaning.
WARNING: Connect only DRYLINE
™
gas sampling lines to the water trap.
Note that there may be other compatible tubes present that
must not be used (e.g. IV lines).
WARNING: Do not use DRYLINE
™
Neonatal sampling lines (blue Luer lock
nuts) with DRYLINE
™
Adult/Pediatric water traps. Do not use
DRYLINE
™
Adult/Pediatric sampling lines (colorless Luer lock
nuts) with DRYLINE
™
Neonatal water traps. These
configurations could result in incorrect measurement data.
WARNING: Do not use Adult/Pediatric type water traps and/or sampling
lines with neonates to avoid high sampling flow.
WARNING: The contents of the Gas Module water trap should be handled
as a potential infection hazard. The water trap, sampling line
and airway adapter should be disposed of in accordance with
local regulations for contaminated and biologically hazardous
items.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...