Passport V Operating Instructions
9 - 1
9.0
IBP Monitoring (optional)
9.1
IBP Description
Intended patient types: adult, pediatric, neonatal.
Invasive Blood Pressure (IBP) is a direct measurement of the patient's arterial or venous blood
pressure. IBP utilizes a catheter that is inserted directly into a vein, artery or other pressure
access areas, and is connected to a transducer for interpretation of Systolic (Sys), Diastolic
(Dia), and Mean blood pressures. Up to two Invasive Blood Pressure channels are available.
Invasive Blood Pressure measurement is an option for the
Passport V
.
The
Passport V
supports the measurements of two invasive blood pressures. It provides
systolic pressure, diastolic pressure and mean pressure values.
When measuring Art/UA/LV/PA pressure, the channel can serve as an HR source.
• When displaying a pulsatile pressure, the format shall be SYS/DIA (MEAN)
• When displaying non-pulsatile pressure, the format shall be MEAN (SYS/DIA)
• When displaying ICP, the format shall be MEAN CePP.
9.1.1
Pressure Labels
IBP1, IBP2 waveforms may be labeled according to the site of insertion. The labels are
identified in the following table:
LABEL
DEFINITION
Art
Arterial Blood Pressure
UA
Umbilical Artery Pressure
LV
Left Ventricular Pressure
PA
Pulmonary Artery Pressure
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...