Alarms
Alarms
13 - 8
Passport V Operating Instructions
“Non-Lethal” Arrhythmia Alarms are priority 2 alarms by default. These alarms will silence
and cancel automatically if the patient's condition is corrected.
E.
ST Alarms
ST alarms can be set for either single lead alarm or dual lead alarm. The ST single lead
alarm will be initiated if one ST lead has exceeded the set alarm parameters. The ST dual
lead alarm will be initiated if two ST leads have exceeded the set alarm parameters.
ST alarms are latching alarms. They will not be silenced unless the operator acknowledges
the alarm by pressing the
Silence
or
Silence All
keys.
F.
Indications of Disabled or Silenced Alarms
ALARM OFF ICON
If any alarm for a monitored parameter is not set, an
Alarm Off Icon
, a triangle with an
“X” through it, will be displayed in the upper area of the screen.
VOLUME -
The audio level of the alarm can be adjusted through the
Monitor
Setup Menu
.
ALARM SILENCE
One or more alarms can be silenced for a programmable length of time. The following is a
description of how to enable the different silence modes.
Silence
- This key silences alarms on parameters for a programmed length of time (default is
2 minutes), or until the alarm condition is no longer present, whichever is shorter. Any new
alarms that occur during the silenced period will disable the silence and the alarm will sound
the tone. An
Alarm Silence Icon
, a bell with an “X” through it, is displayed in the upper
area of the screen. A message and digital timer counts down in the upper message area.
Pressing
Silence
again does not re-enable audio alarms.
Silence All -
This key suspends alarms on all parameters for a programmed period of time
(default is 2 minutes). An
Alarm Silence Icon
, a bell with an “X” through it, is displayed in
the upper area of the screen. A message and timer appear in the upper message area
showing the time remaining. Pressing
Silence All
at any time re-enables audio alarm tones.
If
Silence All For
is set to
Permanent
in the
Alarm Setup
menu, the message
All
Alarms Silenced Permanently
is displayed. Note that the
Permanent
selection will be
available only if
Enable Silence All Permanent Selection
is set to “Yes” in the
Installation Menu
.
The time period for
Silence
and
Silence All
is adjustable via the
Alarm Setup
menu.
13.1.5
Verifying Alarm Functionality
Self-Test of Alarm System
The alarm system will perform a self-test of alarm tones and alarm lights after the
Passport
V
monitor has been powered on.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...