Respiration and CO
2
Troubleshooting
Respiration Monitoring
5 - 6
Passport V Operating Instructions
Resp Disturbed
Patient Movement
External Interference.
Check patient
For impedance - Check
Electrode Contacts / reposition
electrodes/cable
Check environment for source
of interference.
Resp Rate Overrange
Respiration Outside of
measureable range
Check patient.
Resp: Initialization Error
During the Resp module power-
on,as the Resp module
communication stops,system
fails to communicate with
module.
Contact Technical Support.
Resp: Communication Error
System can't communicate
correctly with Resp module,and
receive the error data packets.
Restart the monitor.If the error
still appears,please contact
Technical Support.
Resp: Communication Stop
As the Resp module
communication stops, the data
packets sent by the module
can not be received.
Contact Technical Support.
Resp: High Impedance
Connections not tight and/or
properly secured.
Electrodes dry or loose.
Cable or lead wires damaged.
Ensure proper connection.
(Electrode to lead, lead to
cable, cable to monitor).
Re-prep skin and apply fresh,
moist electrodes.
Check with continuity tester
MESSAGE/PROBLEM
REASON
SOLUTION
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...