IBP Menu
IBP Monitoring (optional)
9 - 6
Passport V Operating Instructions
Scale
-10 to 10 mmHg
0 to 20 mmHg
0 to 40 mmHg
0 to 60 mmHg
0 to 80 mmHg
0 to 140 mmHg
60 to 140 mmHg
0 to 160 mmHg
0 to 225 mmHg
0 to 320 mmHg
Select to change IBP scale.
Defaults:
Art: 0 to 160 mmHg
UA: 0 to 80 mmHg
LV: 0 to 160 mmHg
PA: 0 to 40 mmHg
CVP: 0 to 20 mmHg
ICP: 0 to 20 mmHg
LA: 0 to 20 mmHg
RA: 0 to 20 mmHg
P1, P2: If Measure Pressure is All, default is 0 to
160 mmHg. If Measure Pressure is Mean, default is
0 to 20 mmHg.
IBP Filter Setting
8 Hz
12 Hz
(default)
20 Hz
Select to change IBP filter settings.
The filter setting is linked to both IBP channels.
Zero
—
Select to zero IBP channel.
The IBP transducer must be zeroed before initial
measurements during setup to obtain correct
pressure value. If the transducer is not zeroed the
measured value will be displayed as "XXX".
Implement IBP zeroing when any of the following
occurs:
• Use of new sensor or sensor cables
• Reconnection of the sensor with the monitor
• After monitor restart
• Occurrence of incorrect pressure readings
"Zero ALL" key on the front panel is to zero all IBP
channels. The zero key for a single channel is
available in each IBP menu. The zero key is
disabled until zeroing is completed.
In some conditions, zeroing can not be completed,
and message shall be displayed.
• Pulsatile Pressure. Cannot Zero!
• Pressure Overrange. Cannot Zero!
IBP Menu
MENU ITEM
SELECTIONS
COMMENTS
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...