Passport V Operating Instructions
xv
Precautions
Introduction
CAUTION:
Some disinfectants may cause skin irritation. Please rinse cuff
thoroughly with water to remove any residual disinfectants.
CAUTION:
The internal sampling system of the Gas Module does not need
to be cleaned or sterilized. There is no reverse flow back to the
patient. If the internal sampling system is suspected to be
clogged or dirty, the module should be serviced by an
authorized service person only.
CAUTION:
To avoid permanent damage, do not expose metal components
(pins, sockets, snaps) to disinfectants, soaps or chemicals.
CAUTION:
Do not connect NIBP Luer fittings to Blood Pressure Cuff or
Monitor.
CAUTION:
Some pacemakers may contain a respiratory sensor that may
produce artifact on an ECG waveform.
CAUTION:
The monitor display provides data and waveform information
over its entire area. Do not cover any part of the monitor
display with tape or labels.
CAUTION:
A functional tester cannot be used to assess the accuracy of the
pulse oximeter probe or a pulse oximeter monitor.
CAUTION:
Replace lithium-ion batteries with part number
M05-010001-06 (Mindray) or 0146-00-0099 ONLY.
CAUTION:
Gas Module 3 must be moisture protected whenever
transported. This can be done with a protective plastic bag in
which water-absorbing materials (e.g. silica gel) have been
included.
CAUTION:
Contamination with CO
2
, N
2
O or Anesthetic Agent in the air
surrounding the Gas Module may cause significant
measurement errors. The Gas Module exhaust output should be
connected to a medical gas-scavenging system.
CAUTION:
The Passport V monitor is not compatible with 700 Series
temperature probes.
CAUTION:
Do not unplug the storage device from any of the SB ports on
the Passport V while transferring data, as indicated on the
menu or prompt areas of the monitor. Data may be lost or
corrupted and the storage device may become damaged.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...