Performance Characteristics
Appendix
19 - 16
Passport V Operating Instructions
NIBP Start Pressure Settings and Ranges
The Start Pressure is adjustable throughout the following ranges and is set to the following
defaults:
NIBP Measurement Cycle
The Maximum Measurement Cycle Duration is 180 sec for Adult and Pediatric patients
(Reference: EN60601-2-30, 1995).
The Maximum Measurement Cycle Duration is 90 sec for Neonatal patients (Reference:
EN60601-2-30, 1995).
Pulse Rate Range for NIBP Measurement
19.8.5
Masimo SET SpO
2
Performance Requirements
Sensor Compatibility
LNOP
®
and LNCS
®
Series
Measured Parameters
SpO
2
and PR
SpO
2
Measurement Range
From 1% to 100%
SpO
2
Accuracy
A functional tester or patient simulator not used to assess the accuracy of the pulse oximeter
probe or a pulse oximeter monitor.
PATIENT SIZE
USER-SELECTABLE START
PRESSURE RANGE
INCREMENT SIZE
DEFAULT
SETTING
Adult Mode
100-280 mmHg
5 mmHg
180 mmHg
Pediatric Mode
60-180 mmHg
5 mmHg
140 mmHg
Neonatal Mode
40-120 mmHg
5 mmHg
100 mmHg
PATIENT SIZE
RANGE
ACCURACY
Adult Mode
35 - 245 bpm
The greater of ± 3 bpm or ± 3%
Pediatric Mode
35 - 245 bpm
The greater of ± 3 bpm or ± 3%
Neonatal Mode
70 - 245 bpm
The greater of ± 3 bpm or ± 3%
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...