Prompt Messages
Messages
14 - 22
Passport V Operating Instructions
14.3.15
Alarm Troubleshooting
MESSAGE/PROBLEM
REASON
SOLUTION
High or Low or No Alarm
Sound
Alarm limits not set
Go to
Alarm Setup
and adjust
alarms.
Alarm
Silence All, On
time has
not expired
Press
SILENCE ALL
to reactivate
alarms.
No Arrhythmia
Alarm Sound
Arrhythmia option not installed
Call Sales Rep to purchase
option.
Arrhythmia Alarms off
Go to
Monitor Setup
〉
Advanced Setup
to activate
alarm.
Monitor is in learning mode
Wait until learning is concluded
and monitor patient closely.
Alarms continue to Sound
despite pressing
SILENCE
More than one alarm is active
Press
SILENCE
or
SILENCE ALL
key to silence.
Check Patient.
No Alarm printout with
Alarm violation
Print on Alarm
is set to
Suspend.
Go to
Print Menu
and set
on Alarm
to
Active
.