Passport V Operating Instructions
3 - 7
Patient Management
Remote View
All Remote View controls are contained within the
Remote View Menu
. Opening other
menus from the monitor’s keypad, or pressing the
Normal Screen
key, will close the
Remote View Menu
.
The
Remote View Menu
displays data as follows:
• The waveforms that are displayed are user-selectable through the
Remote View
Setup Menu
, as described in the table on page 3-8. Corresponding numeric tiles
are displayed to the right of the waveforms. The numeric data for a maximum of fifteen
(15) parameters that are not related to the waveforms is displayed in the “Numeric
Data Area” located directly below the waveforms.
• The colors of the waveforms and all numeric data correspond to the settings of the host
monitor (not the remote monitor).
• When numeric data for a particular parameter is unavailable, dashes (– – –) are
displayed.
• If a high or low alarm limit has not been set, an Alarm Off icon is displayed.
• Remote View alarms follow the alarm settings at the remote monitor, displaying in a
priority-appropriate, reverse-video color.
• If the remote monitor is placed into Standby mode, the message “In Standby” is
displayed in the Remote View Window.
NOTE:
Respiration alarm text messages are also displayed in the
displayed in the message area (upper part) of the Remote
View Menu window as follows:
The
Remote View Menu
also provides the following menu choices:
Normal Screen
,
Select Beds
,
Setup
,
Remote Silence
,
Remote Silence All
,
and
Remote Print
.
1. Normal Screen
— this selection removes the
Remote View Menu
from the display.
2. Select Beds
— this selection opens the
Select Beds
menu from which beds may be
selected to be included in the care group. All networked beds will be in que but only 8
can be selected.
• If Apnea is detected, the message “APNEA” is displayed.
• If CVA is detected, the message “CVA” is displayed.
• If high impedance is detected from the ECG electrodes, the
message “CHK LEAD” is displayed.
NOTE:
Remote View function is not available between wireless
monitors, nor for wireless monitors to view wired monitors.
Remote View function is available between wired monitors,
or for wired monitors to view wireless monitors.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...