Passport V Operating Instructions
4 - 9
ECG Monitoring
Menus: ECG Main and Submenus
To display the Arrhythmia Menu:
1.
On the front panel, press the
Alarms: Limits
key to display the
Alarm Settings
Menu
, and select
Arrhythmia Menu
or
On the normal screen:
Select
Monitor Setup
〉
Advanced Setup
〉
Arrhythmia Menu
or
Select
Parameters
〉
ECG
〉
Arrhythmia Menu
or
Select the
ECG
tile
〉
Arrhythmia Menu
.
2.
To exit, press the
Normal Screen
key on the front panel.
Arrhythmia Menu
MENU ITEM
SELECTIONS
COMMENTS
Previous Menu
—
Select to return to previous menu.
Arrhythmia Setup
Asystole
V-Tach
V-Fib
PVC/min
V-Rhythm
Couplet
Run
Bigeminy
Trigeminy
Irregular HR
Brady
Select to set individual arrhythmia alarm
characteristics.
For each arrhythmia selection, choose:
Alarm: On or Off
Priority: 1 or 2
Print: On or Off
Lethal arrhythmias (Asystole, V-Tach, and V-Fib) are
Priority 1 and cannot be changed. The factory default
for all arrhythmia alarms is On.
All On
—
Select to set all arrhythmia alarms to On.
Non-lethals off
—
Select to set all non-lethal arrhythmia alarms to Off.
All Off
—
Select to set all arrhythmia alarms to Off.
“All Off” is displayed only if “Enable Arrhythmia All
Off Selection” = Yes in the
Installation Menu
.
Relearn
—
Select to manually initiate the relearning process for ST
Measurements or Arrhythmia Analysis.
Disabled if Arrhythmia = Off.
After relearning is started, arrhythmia relearning and
ST relearning are performed simultaneously. The
arrhythmia and ST analysis templates are recreated
and the message “Relearning...” is displayed. During
relearning, ST and PVCs are displayed as “– – –“.
Arrhythmia relearning should be started when:
• Lead type is changed.
• ECG lead wires are reconnected to the patient.
• Analysis lead is changed.
• Patient size is changed.
• Arrhythmia analysis is turned on.
• Module is turned on.
• Exiting Standby mode.
PVC/min
1 to 30
default = 10
Select to change the High PVC/min alarm rate limit.
The PVC/min parameter is displayed when PVC/min
arrhythmia alarm is set to On.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...