Passport V Operating Instructions
8 - 3
Temperature Monitoring
Temperature Probes
8.3
Temperature Probes
CAUTION:
The Passport V monitor is not compatible with 700 Series
temperature probes.
8.3.1
Skin Temperature Sensor with 400 Series Thermistor
Description (0206-03-0300-02)
Monitoring of skin surface temperature is a well accepted procedure for the detection of
hypothermic and hyper thermic conditions. The Skin Temperature Sensor is designed for
placement on the surface of the skin. The skin temperature sensor consists of a thermistor
embedded in an adhesive backed foam disk. One side of the foam disk is covered by a
metallic film. The opposite surface of the foam is coated with a medical grade hypo-
allergenic adhesive suitable for skin application. The hypo-allergenic adhesive holds the
sensor in situ. The foam thermally insulates the sensor and the metallic layer reflects external
infrared heat energy thus providing accurate measurement of body temperature.
The temperature transducer is in the form of an electrically insulated 400 series thermistor.
The accuracy of the thermistor is ± 0.2° Celsius in the range of 5° Celsius to 45° Celsius. The
assembly is disposable, single use only, and packaged sterile.
This device is intended for use with electronic temperature monitors accepting 400 series
thermistor sensors or equivalent that are equipped with alarms. To interconnect the probe
with the instrument, use cable part number 0012-00-0975.
Electrical leakage current of the device, when used with the monitor and cable, comply with
IEC 601-1/EN60601-1.
Indications
The Skin Temperature Sensor is indicated for use in the routine monitoring of skin
temperature.
Contraindications
The Skin Temperature Sensor may be contraindicated over traumatized areas.
Adverse Reactions
Adverse reactions reported during applications of Skin Temperature Sensors include: skin
abrasions and tissue burns due to aberrant electro-cautery current pathways.
Directions For Use And Precautions
1.
Dry skin completely in the area intended for sensor placement.
2.
Remove the protective backing and place the sensor onto the previously dried skin area.
3.
Align the sensor’s connector with the monitor cable’s connector and push firmly to assure
full contact. Forced mating of the connectors without proper alignment may cause
damage to the connectors and interruption in electrical continuity.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...