Passport V Operating Instructions
10 - 5
CO
2
Monitoring (optional)
CO
2
Menu
10.3.1
CO
2
Setup Menu (DPM Sidestream Only)
To display the
CO
2
Setup Menu
:
1.
On the front panel:
Press the
Normal Screen
key to return to the normal screen.
On the normal screen:
Select
Parameters
〉
CO
2
〉
CO
2
Setup
or
Select the
CO
2
tile
〉
CO
2
Setup
.
NOTE:
To extend the lifetime of the watertrap and module,
disconnect the watertrap and set the operating mode to
Standby mode when CO
2
monitoring is not required.
NOTE:
The watertrap collects water drops condensed in the
sampling line and therefore prevents them from entering
the module. It is important to periodically drain the
watertrap to avoid water blocking the airway.
Selection
Gas Module 3
Status
Internal CO
2
Module Status
Active CO
2
Source
Gas Module
OFF
OFF
None
Gas Module
OFF
ON
Internal CO
2
Module
Gas Module
ON
OFF
Gas Module
Gas Module
ON
ON
Gas Module
Internal CO
2
Module
OFF
OFF
None
Internal CO
2
Module
OFF
ON
Internal CO
2
Module
Internal CO
2
Module
ON
OFF
Gas Module
Internal CO
2
Module
ON
ON
Internal CO
2
Module
Note
: ON/OFF Status of the Module:
If the Gas Module is connected to a serial port configured to ‘Gas Module’, and
the power is switched on, the module is considered to be ON. Otherwise it is
considered to be OFF.
If the Oridion CO
2
Module is installed and a Filterline is connected, the module
is considered to be ON. Otherwise it is considered to be OFF.
If the DPM CO
2
Module and water trap are installed, the module is considered
to be ON, Otherwise it is considered to be OFF.
Flow Rate
70 ml/min
100 ml/min
(default)
120 ml/min
150 ml/min
Applicable to DPM Sidestream CO
2
module only.
Note
: For Oridion Microstream CO
2
, the
flow rate is 50ml/min. It is not user
adjustable.
CO
2
Menu
MENU ITEM
SELECTIONS
COMMENTS
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...