Passport V Operating Instructions
7 - 13
NIBP Monitoring
NIBP Troubleshooting
7.5
NIBP Troubleshooting
Refer to Chapter 14.0 Messages for Physiological Alarm Messages, Technical Alarm
Messages, and Prompt Messages.
NOTE:
Always have an alternate method of BP verification
available.
NOTE:
On vasoconstricted patients, failure to evacuate air from the
cuff can distort BP measurement.
NOTE:
Cuff should be at heart level.
NOTE:
The presence of an arrhythmia may increase the time
required to complete a measurement and may cause
measurement cycle to extend beyond a point where a
measurement can be completed.
MESSAGE/PROBLEM
REASON
SOLUTION
Unable to obtain a BP
Patient movement
Wait until patient is calm or
gently hold limb.
Cuff or hose NOT attached /
leaking
Check all connections.
HR irregular / arrhythmia present
Check Patient and notify
Physician.
Blood pressure is out
of range.
Check Patient and verify BP
with manual method.
Improper cuff size / brand
Measure patient limb. Use
only properly sized approved
accessories.
Reading too high or too low
Incorrect cuff size
Measure Patient limb, use
correct cuff.
Patient movement
Wait until patient is calm or
gently hold limb.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...