IBP Troubleshooting
IBP Monitoring (optional)
9 - 8
Passport V Operating Instructions
9.5
IBP Troubleshooting
MESSAGE/PROBLEM
REASON
SOLUTION
Damped Invasive
Waveform
Air bubbles in tubing.
Eliminate air from tubing.
Kinked catheter.
Change position of catheter,
check patient.
Catheter against wall of blood
vessel.
Check for leaks at connector, flush
catheter.
Blood in tubing.
Pump pressure bag up to 300
mmHg.
Catheter partially occluded with
clot.
Consult physician.
IBP not Displayed / No
IBP Waveform
Improper Setup.
Check display setup in monitor
setup.
Cable not plugged in.
Check cable.
Transducer not connected.
Check transducer connection.
Stopcock turned improperly.
Check transducer.
Transducer not zeroed.
Check and zero the transducer.
Dashes are displayed
The measured result is invalid or
out of range.
IBP might be set to non-pulsatile
labels like CVP, LA, RA, and ICP.
Abnormally High or Low
readings
Transducer too HIGH or to LOW.
Check patient adjust transducer
rezero.
Unable to Zero
Stopcock not open to
atmosphere.
Check transducer.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...