Passport V Operating Instructions
4 - 19
ECG Monitoring
Preparation and Lead Placement
4.5.3.5
Lead Placement: Lead II Monitoring
The recommended lead placement for Lead II monitoring is as follows.
FIGURE 4-15
Lead II Monitoring (AHA)
FIGURE 4-16
Lead II Monitoring (IEC)
• Place the RA (white) electrode under the
patient’s right clavicle, at the mid-
clavicular line within the rib cage frame.
• Place the LA (black) electrode under the
patient’s left clavicle, at the mid-
clavicular line within the rib cage frame.
• Place the LL (red) electrode on the
patient’s lower left abdomen within the
rib cage frame.
Select ECG Lead II on the monitor. Lead II is
the direct electrical line between the RA
(white) electrode and the LL (red) electrode.
• Place the R (red) electrode under the
patient’s right clavicle, at the mid-
clavicular line within the rib cage frame.
• Place the L (yellow) electrode under the
patient’s left clavicle, at the mid-
clavicular line within the rib cage frame.
• Place the F (green) electrode on the
patient’s lower left abdomen within the
rib cage frame.
Select ECG Lead II on the monitor. Lead II is
the direct electrical line between the R (red)
electrode and the F (green) electrode.
RA
LA
LL
Black
White
Red
R
L
F
Red
Green
Yellow
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...