Passport V Operating Instructions
11 - 7
Gas Monitoring (optional)
Gas Menu
NOTE:
The Gas Module 3 uses a fixed correction of 11 hPa to
compensate for the influence of water vapor in the gas
sample, when converting the gas readings to ATPD. An
increase in the ambient H
2
O partial pressure to 30 hPa
(i.e. 28 °C, 80% RH or 33 °C, and 60% RH) will cause a
general error for all gases of only -2% REL.
Monitoring Anesthetic Gases, O
2
, N
2
O and/or CO
2
NOTE:
To prevent moisture from entering the pneumatic system,
ensure that the Gas Module is always installed and
operated in the horizontal orientation shown in all graphical
depictions.
1.
Power ON the Gas Module.
2.
Power ON the patient monitor while holding down the
Discharge
key to enter into its
Installation Mode.
3.
From the patient monitor
Installation Menu
, set the correct serial port to “Gas
Module” and then select “Save Current”.
4.
Restart the patient monitor
to enter normal monitoring mode.
5.
Set alarms as desired.
NOTE:
DRYLINE
™
Sample Lines are for use with Gas Module 3 only.
6.
For non-intubated patients, apply the nasal cannula to the patient. For intubated patients
connect the sample line to the breathing circuit. Refer to instruction provided in the
sample line packets.
7.
Connect the other end of the nasal cannula or sample line to the Gas Module at the
input port.
WARNING: Connection of the Gas Module exhaust port to the hospital’s
CAUTION:
8.
Check for a clean water trap.
9.
Observe the capnogram on the monitor’s display. On
Passport V
powerup, O
2
, Agent
and N
2
O numbers will display. CO
2
numbers will be displayed when a valid breath is
detected.
NOTE:
The Gas Module 3 must be warmed up a minimum of 45
seconds for ISO accurate CO
2
, O
2
, N
2
O, and agent readings.
NOTE:
Consideration should be given to the disposal of packaging
waste.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...