Performance Characteristics
Appendix
19 - 8
Passport V Operating Instructions
Common Mode Rejection Ratio
• Monitor mode: >105 dB measured at 50 and 60 Hz
• Surgical mode: >105 dB measured at 50 and 60 Hz
• ST mode: > 105 dB measured at 50 and 60 Hz
• Diagnostic mode: >90 dB measured at 50 and 60 Hz
Notch Filter
• Notch filter settable to 50 Hz or 60 Hz
Input Bias Current
•
≤
0.1 µA .
• Complies with 4.2.5 of ANSI/AAMI EC13:2002.
Electrode Offset Potential Tolerance
• +/- 500 mV
• Complies with 4.2.9.1 of ANSI/AAMI EC13:2002
RTI Noise
•
≤
30 µV (p-v RTI)
• Complies with 4.2.9.3 of ANSI/AAMI EC13:2002.
ESU (Electrosurgical Unit) Interference Suppression
• Incision mode: 300W
• Coagulation mode: 100W
• Recovery time :
≤
10s
• Complies with 4.2.9.14 of ANSI/AAMI EC13:2002.
RTI Suppression of ESU (Electrosurgical Unit)
• ESIS ECG cable is used.
• ECG lead wires complying with AAMI is used; compared with ECG baseline, the peak-to-
peak RTI value:
≤
2mV.
• The testing method complies with 5.2.9.14 of EC13:2002.
ESU Filter Control
• Auto / Off indications:
• Auto: ESU detection shall engage the surgical mode filter if the ESU detection
measures ESU pulses at a rate > 50 Hz or > 50 spikes per second. ESU detection shall
stop the surgical mode filter if the ESU detection measures ESU pulses at a rate < 10
Hz or < 10 spikes per second.
• Off: ESU filter is off.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...