Performance Characteristics
Appendix
19 - 22
Passport V Operating Instructions
miniMediCO
2
Microstream
®
Accuracy specification of the measured CO
2
partial pressure is according to the
following table (This testing is done according to ISO 21647 clauses 51.101.1 and
51.101.2.).
Accuracy in the presence of interfering gases:
The accuracy specification is maintained to within 4% of the values indicated in the table
above in the presence of interfering gases according to ISO 21647 clauses 51.101.3 and
101.1. The accuracy specification is maintained to within 4% of the values indicated in the
table above in the presence of up to 80% Helium with up to 15% oxygen when tested
according to the procedures defined by ISO 21647.
Respiration Rate Accuracy:
Self Maintenance (SFM) Interval:
Self-Maintenance (SFM) is performed only during measurement mode. The module performs
one or more of the following:
• Ambient pressure measurement
• Auto zero (AZ)
• Flow test
SFM is triggered under the following conditions:
• During the first hour after entering measurement mode, periodically for durations of
typically 10 seconds at a rate which limits the total time consumed by SFMs to less
than 2% of the time in which active measurements are taken. Following the first hour
Range:
0 - 99 mmHg
CO
2
PARTIAL PRESSURE *
ACCURACY**
0 - 38 mmHg
± 2 mmHg
39 - 99 mmHg
± (5% of r 0.08% for every 1 mmHg above 38 mmHg)
*
At sea level.
**
Accuracy applies for respiration rates of up to 80 RPM. For respiration rates above 80 RPM, accuracy is
4 mmHg or
±
12% of the reading, whichever is greater, for EtCO
2
values exceeding 18 mmHg. This is
tested according to and is compliant with ISO 21647. To achieve the specified accuracies for respiration
rates above 60 respirations/minute, the Microstream FilterLine H Set for Infant/Neonatal
(p/n 006324) must be used. Above 55 degrees C module temperature,
±
1 mmHg or
±
2.5% (whichever is
greater) has to be added to the tolerance of the accuracy specs.
Flow Rate:
50 ml/min -7.5 + 15 ml/min, flow measured by volume.
Respiration Rate Range:
0 - 150 RPM
Respiration Rate
0 to 70 RPM
71 to 120 RPM
121 to 150 RPM
Accuracy
± 1 RPM
± 2 RPM
± 3 RPM
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...