Passport V Operating Instructions
14 - 7
Messages
Technical Alarm Messages
14.2.9
System Level
ALARM MESSAGE
PRIORITY
REASON
SOLUTION
Low Battery
3
The battery capacity
is lower than the set
alarm limit value.
Change the battery, or use the
AC mains.
12V Too High
1
The monitoring of the
internal voltage finds
that 12V voltage is
too high.
Contact Technical Support.
12V Too Low
1
The monitoring of the
internal voltage finds
that 12V voltage is
too low.
Contact Technical Support.
5V Too High
1
The monitoring of the
internal voltage finds
that 5V voltage is too
high.
Contact Technical Support.
5V Too Low
1
The monitoring of the
internal voltage finds
that 5V voltage is too
low.
Contact Technical Support.
3.3V Too High
1
The monitoring of the
internal voltage finds
that 3.3V voltage is
too high.
Contact Technical Support.
3.3V Too Low
1
The monitoring of the
internal voltage finds
that 3.3V voltage is
too low.
Contact Technical Support.