Passport V Operating Instructions
19 - 15
Appendix
Performance Characteristics
Maximum Cuff Pressure
Cuff Inflation
The
Passport V
is capable of supplying sufficient air to bring a volume of at least 200 cc
(12 cubic inches) to a pressure of 300 mmHg in no more than 10 seconds, unless otherwise
stated. (ANSI/AAMI SP10A-1996)
If the cuff is not inflated to the desired pressure within 60 seconds then the cuff is vented and
a retry cycle is initiated, up to 3 times.
Maximum Leakage
The maximum allowed pressure drop with the bleed valves closed is 10 mmHg in 10 seconds
as measured with a 500 cc volume at differential pressures of 250 mmHg, 150 mmHg and
50 mmHg (Reference ANSI/AAMI SP10-1992, EN1060-3, 1997).
Vent Rate
A volume of 500 cc, when vented, is reduced from a pressure of 260 mmHg to a pressure of
15 mmHg in a maximum of 10 seconds. For Neonate, from a pressure of 150 mmHg to a
pressure of 5 mmHg in less then 5 seconds (Reference: EN1060-3, 1997).
Initial Conditions
An NIBP Zero is performed automatically before the NIBP can be initiated.
Hardware Maximum Cuff pressure:
Adult Mode:
330 mmHg
Pediatric Mode:
220 mmHg
Neonatal Mode:
165 mmHg
Software Maximum Cuff pressure:
Adult Mode:
300 mmHg
Pediatric Mode:
200 mmHg
Neonatal Mode:
145 mmHg
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...