Passport V Operating Instructions
4 - 33
ECG Monitoring
Relearning ST or Arrhythmia Analysis
4.10
Relearning ST or Arrhythmia Analysis
Automatic Relearning
The
Passport V
initiates the learning process for ST measurements or Arrhythmia analysis
after one of the following:
• Unit Power-Up
• Return to normal monitoring from
Standby
mode
• Enabling ST or Arrhythmia analysis
• The lead has been changed in ECG 1 waveform (3-lead only)
•
Patient Size
is changed
• Whenever the
Relearn
function is selected from the
ST
,
ECG
or
Arrhythmia
Menus
• The monitor switches between 3-lead and 5-lead mode
Manual Relearning
To initiate a manual relearn, select
Relearn
from the
ECG Menu
,
Arrhythmia Menu
, or
ST Menu
.
It is recommended to initiate a manual Relearn after one or more of the following:
• The ECG electrodes have been repositioned
• Sufficient time has passed since the last Relearn
• After significant changes to the patient QRS complex
• After significant changes to the patient ECG rhythm
• A clinician has observed clinically questionable arrhythmia calls
A Relearn must be initiated if “Learning” occurred during a “Leads Off” condition.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...