Graphic Trends
Trends
15 - 6
Passport V Operating Instructions
As the cursor is scrolled horizontally, the digital data corresponding to the points in the graph
is shown at the right side of the window. Trend data in violation of an alarm is highlighted
according to the priority of the alarm.
On color displays:
• Priority 1 alarm data is shown in inverse video with red background.
• Priority 2 alarm data is shown in inverse video with yellow background.
If data for a parameter is not available at the time of the trend entry, the data field will be
dashed. If an NIBP reading could not be obtained or an invasive pressure channel was not
zeroed at the time of the trend entry, the data field will contain
“XXX”
.
Modification of Parameters Displayed
The parameters displayed always include the currently active parameters and any others
used since the time the patient was admitted to the monitoring system. The default order of
parameters displayed from top to bottom is: HR, SpO
2
, NIBP, Resp, CO
2
, O
2
, Des, N
2
O,
MAC, P1, P2, Temp, and PVC. To change the order in which parameters are displayed in
each of the top 5 rows, select
Group Setup
from the
Graphic Trend
menu. Then, set
Param 1
through
Param 6
to the desired parameters.
Modification of Trend Entry Conditions
The
Graphic Trend
data is the same as that stored for
List Trends
, arranged graphically.
(If data is not available for time period, it appears as a gap in the
Graphic Trend
.) Refer to
the previous section (
List Trend
) for modification of trend entry conditions.
Transferring Graphic Trend Data Between Different
Passport V
Monitors
Graphic Trend
data is transferred together with
List Trend
and patient name and
demographics. Refer to the
List Trend
section for details.
Clearing Trend Data
To manually clear all trend data, including List trends, choose
Clear Trends
from the menu.
A confirmation prompt will appear. Once cleared, the data cannot be restored.
• All trend data is automatically cleared when the patient is discharged from the
monitor.
• If the monitor’s displayed time or date is changed, an event will be marked but not
cleared of all trend data.
Removing the Graphic Trend Display
The
Graphic Trend
display does not automatically time-out and must be manually removed
to return to the normal waveform display. To remove the Graphic Trend display, choose
Normal Screen
from the menu, or press the
Normal Screen
key.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...