Keys and Front Panel
General Product Description
1 - 6
Passport V Operating Instructions
13. Print Trend
Press this key to initiate printing of the displayed trend. By default, the monitor’s stored trend
information will be printed by the internal printer. Pressing this key during a print job will
abort printing.
• If the print destination is a remote Central Station, then pressing this key will initiate a
trend report at the Central Station.
• If the print destination is a remote printer, then pressing this key will initiate a trend report
at the remote printer.
14. Standby
Press this key to place the
Passport V
into a STANDBY mode. While in STANDBY mode,
monitoring is discontinued and the alarms are in permanent suspension, interval NIBP
measurements are placed in idle mode, CO
2
pump is shut off, and the display shuts down.
The monitor notifies the central station and other monitors.
Except for the power on/off switch
and the STANDBY key, all the other hard keys on the front panel are disabled. When in the
STANDBY mode, the message
To Begin Monitoring, press Standby.
is displayed.
NOTE:
Trend data is not cleared in the STANDBY mode. When the
STANDBY mode is released, NIBP INTERVAL is in IDLE MODE
and requires reactivation via the START key. The CO
2
pump
automatically reactivates if the DPM or Microstream
®
sensor
is in place.
Press the
Standby
key to exit the STANDBY mode and return to the normal screen. When
exiting from the standby mode to the monitoring mode, the monitor responds as follows:
• Restores parameter measurement: Start arrhythmia relearning. Gas modules recover the
working status before the standby mode. NIBP measurement will not start automatically.
• Restores normal data storage.
• All the alarms are activated.
• The screen becomes the one displayed before the standby mode.
• The function of auxiliary output is restored.
• The functions of input devices (such as the hard keys on the front panel) are activated.
• The monitor notifies the central station and other monitors.
15. Discharge
Press this key to discharge a patient.
16. Mark Event
Press this key to cause a time stamp event marker to be noted in the trend memory.
17. Trends
Press this key repeatedly to toggle through the following screens:
Quick Trend
,
List Trend
,
Graphic Trend
, and normal screen (or
OxyCRG
, if the patient size is set to Neonate).
18. Normal Screen
Press this key at any time to return the screen to the normal monitoring mode. All menus are
closed.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...