Prompt Messages
Messages
14 - 12
Passport V Operating Instructions
NIBP: STAT
Displayed while the Interval is
set to STAT.
Press START to initiate a STAT
measurement sequence or use
the ‘Interval’ key to change the
setting to Off or a timed
interval value.
NIBP: Warming Up
Warming Up
This is normal operation. Wait
for message to clear.
NIBP: Retry Pump Higher
A measurement has been
attempted but no reading was
possible. This results from
inadequate cuff inflation
pressure.
Retry will be attempted.
Check that appropriate patient
size is set.
Preset initial inflation
pressure.
NIBP: Retry
Cuff is loosely wrapped. The cuff
may be too loosely wrapped or
not attached at all.
Weak signal. The pulse of the
patient may be too weak or the
cuff is loosely wrapped.
Excessive motion. In
measurement, signals contain
motion artifact or too much
interference.
Timeout. Measurement takes
more than 120 seconds in ADU/
PED mode and 90 seconds in
NEO mode.
Retry Overpressure or bad
measurement.
Retry will be attempted. Check
for leaks and quality of
peripheral pulses. Decrease
patient movement. Switch cuff
to another limb.
Accuracy Testing...
Appears when press Accuracy
Test button in Calibrate NIBP
Menu.
/
Accuracy Testing Complete.
During the accuracy testing,
press Stop Accuracy Test button
in Calibrate NIBP Menu, this
message will display.
/
Leakage Testing...
Appears when press Leakage
Test button in Calibrate NIBP
Menu.
/
Leakage Testing Complete
During the leakage testing, press
Stop Leakage Test button in
Calibrate NIBP Menu, this
message will display.
/
Calibrating...
Occurs when NIBP module is
calibrating.
Wait until calibration is
completed
Calibration Failed
During calibration, departure
occurs between the
measurement pressure and
reference pressure due to
improper operation or
measurement error.
Repeat calibration
procedure.
Содержание Passport V
Страница 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Страница 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Страница 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 177: ...Alarms Alarms 13 10 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Страница 286: ...This page intentionally left blank ...
Страница 287: ...0070 00 0704 02 Rev 13 0 March 2015 ...