3.2 Configuring Measurement Settings
71
3
3
Chapt
er
3 Sett
ings
Example: If the recording interval is set to 100 ms while using the LR8510
and LR8513
Data update interval
Data is updated at the unit's minimum recording interval, regardless of the
instrument's recording interval setting.
Unit
Minimum recording interval
(data refresh interval)
LR8510
100 ms
LR8511
100 ms
LR8512
100 ms
LR8513
500 ms
LR8514
500 ms
LR8515
100 ms
LR8520
500 ms
If the recording interval setting is shorter than the data update interval, there
will be some intervals when the data is not updated. Since the measured val-
ue shown at such times will be the same as the value for the last data up-
date, the waveform will take on a stepped appearance.
LR8510
LR8513
100 ms
(data refresh interval)
500 ms
(data refresh interval)