Appendix 1 Scan Timing
A
1
Appendix
L
For sampling of the LR8510 Wireless Voltage/Temp Unit and LR8511 Wireless Universal
Unit, there is a scan system for each channel.
The following figure illustrates scan timing when measurement is on for all 15 unit chan-
nels with a 50/60 Hz digital filter and recording intervals of 100 ms and 5s, respectively.
When the digital filter is off, the timing is the same as with a recording interval of 100 ms,
regardless of the recording interval setting.
When the digital filter is set to 50 Hz or 60 Hz, the scan width varies with the recording
interval. If the recording interval is 5 s, the scan width per channel will be approximately
120 ms. Even with a recording interval that is slower than 5 s, the scan width per channel
will be set at approximately 120 ms.
The maximum disparity in sampling among measurement units is 1 sample (recording
interval), regardless of the number of units.
Appendix
Appendix 1 Scan Timing
Recording interval:100 ms
Approx. 5 ms
CH1
CH2
CH3
CH14 CH15
CH1
CH2
Recording interval: 5 s
CH1 CH2
CH14 CH15
CH1 CH2
Approx. 120 ms
CH3
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......