3.3 Input Channel Settings
81
3
3
Chapt
er
3 Sett
ings
Make these channel settings for resistance measurement.
Make these settings on the
[CH]
screen.
See:
"Key Setting Procedure" (p .73)
Resistance Measurement Settings
When measuring an inductive load, for example from a wire-wound resis-
tor, the instrument’s response may not be fast enough when the inductor
value is 50 mH or greater, preventing accurate measurement.
1
Select the input unit (
UNIT1
to
7
) and channel
(
1-1
to
7-15
), and check the box to enable the
channel.
2
Select
[Resist]
.
3
Select the appropriate measurement ampli-
tude range.
Setting options:(
: default setting)
10
, 20
, 100
, 200
: On
: Off
1
2
3
When a measured value far exceeds the range limit (range label)
"OVER" or "-OVER" is displayed in place of numerical/cursor values, and the following values
are applied to calculations and data storage.
Input type
Input range +OVER value -OVER value
Resistance
10
16.3835
-16.384
20
32.767
-32.768
100
163.835
-163.84
200
327.67
-327.68
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......